| Model Number |
METRIX AIR |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Codes |
Dizziness (2194); Polydipsia (2604)
|
| Date of Event |
09/28/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Product not returned for evaluation.No replacement product at this time.Test strips are expired.Most likely underlying root cause: mlc-22-client is testing with expired test strips test strip udi# (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call in order to ensure the customer's condition since the initial call - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for e-0 accompanied by symptoms.The expected fasting blood glucose test result range is undisclosed.Customer did report symptoms of feeling light headed and dry mouth.Medical attention is not reported as a result of the actual blood glucose results and reported symptoms.The product storage location is undisclosed.Test strip lot manufacturer's expiration date is 08/29/2017 (expired) and open vial date is undisclosed.The meter memory was not reviewed for previous test result history.
|
| |
|