| Model Number |
1882940HR |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/24/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
A health care provider (hcp) reported that the blade had came apart once it was used during functional endoscopic sinus surgery (fess).The o-ring came off completely and got stuck in the handpiece but it was also retrieved.There was no delay in the surgical procedure and a back-up device was used and it functioned fine.There was no patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis found that visually, the inner green seal was in place and in good condition when received.Functionally, the blade was loaded in and out of a handpiece 10 times and the seal did not come out of its seated position.There are no signs of a defect with the device.If information is provided in the future, a supplemental report will be issued.
|
| |
|