Model Number 1882940HR
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/24/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A health care provider (hcp) reported that the blade had came apart once it was used during functional endoscopic sinus surgery (fess).The o-ring came off completely and got stuck in the handpiece but it was also retrieved.There was no delay in the surgical procedure and a back-up device was used and it functioned fine.There was no patient impact.
 
Additional Manufacturer Narrative
Analysis found that visually, the inner green seal was in place and in good condition when received.Functionally, the blade was loaded in and out of a handpiece 10 times and the seal did not come out of its seated position.There are no signs of a defect with the device.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XPS® BLADE
Common Device Name
BUR, EAR, NOSE AND THROAT
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
MDR Report Key7996638
Report Number1045254-2018-00535
Device Sequence Number11109133
Product Code EQJ
UDI-Device Identifier00885074079813
UDI-Public00885074079813
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 11/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/24/2022
Device Model Number1882940HR
Device Catalogue Number1882940HR
Device Lot NumberHG2PNH2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/29/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/28/2018
Supplement Date Received by Manufacturer10/30/2018
Initial Report FDA Received Date10/23/2018
Supplement Report FDA Received Date11/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age25 YR
Patient Weight86
-
-