| Model Number |
CRPLUS |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
10/04/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Physio-control continues to investigate the reported issue and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
| |
|
Event or Problem Description
|
|
A customer contacted physio-control to report that their device was showing all three icons (attention, charge-pak and service wrench).With all three icons illuminated, the device may not have sufficient power available to deliver effective defibrillation therapy to a patient, if it was needed.There was no patient use associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Physio-control evaluated the customer's device and could not duplicate the reported issue.The customer did not return the charge-pak with the device.After observing proper device operation through functional and performance testing the unit was returned to the customer for use.The cause of the reported issue could not be conclusively determined.
|
| |
|
Event or Problem Description
|
|
A customer contacted physio-control to report that their device was showing all three icons (attention, charge-pak and service wrench).With all three icons illuminated, the device may not have sufficient power available to deliver effective defibrillation therapy to a patient, if it was needed.There was no patient use associated with the reported event.
|
| |
|