• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL ULTIMUM EV HEMOSTASIS INTRODUCER WITH TIP MARKER; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL ULTIMUM EV HEMOSTASIS INTRODUCER WITH TIP MARKER; INTRODUCER, CATHETER Back to Search Results
Model Number C407699
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/17/2018
Type of Reportable Event Malfunction
Event or Problem Description
During the portico procedure, the valve of the ultimum ev hemostasis introducer was leaking blood.The portico valve was implanted without any complications and the procedure was completed without using another ultimum ev hemostasis introducer.Patient's information (e.G.Age, weight, gender, ethnicity, race, patient's initials) cannot be handled by abbott in absence of patient's written consent as required by the personal data protection national legislation.National legislation prevents the recording of such information.A written consent has not been obtained in this case; therefore, this information is not available.
 
Additional Manufacturer Narrative
The reported event of a leak could not be confirmed.The sheath passed pressure and aspiration leak testing with no anomalies observed.However, microscopic inspection revealed tearing on the proximal face of the seal.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported event remains unknown.The cause of the torn seal is consistent with damage during use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTIMUM EV HEMOSTASIS INTRODUCER WITH TIP MARKER
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
MDR Report Key7996787
Report Number3005334138-2018-00383
Device Sequence Number1386675
Product Code DYB
UDI-Device Identifier05415067014382
UDI-Public05415067014382
Combination Product (Y/N)N
Initial Reporter CountryMX
PMA/510(K) Number
K140327
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/29/2020
Device Model NumberC407699
Device Catalogue NumberC407698
Device Lot Number6381038
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 10/17/2018
Supplement Date Received by Manufacturer02/07/2019
Initial Report FDA Received Date10/23/2018
Supplement Report FDA Received Date02/07/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-