BECTON DICKINSON MEDICAL SYSTEMS BD LUER-LOK¿ SYRINGE STERILE, SINGLE USE; PISTON SYRINGE
|
Back to Search Results |
|
Catalog Number 309646 |
Device Problem
Difficult to Remove (1528)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/03/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
|
Event Description
|
It was reported that a bd luer-lok¿ syringe sterile, single use had a plunger that was hard to pull back.There was no report of exposure, serious injury, or medical intervention.
|
|
Manufacturer Narrative
|
Investigation summary: since no samples displaying the condition reported are available for examination, we were unable to fully investigate this incident, therefore a root cause could not be determined.Dhr review for batch 8035612 ((b)(4).Manufacturing dates: 02/12/2018 to 02/14/2018.Batch quantity was 480,000.All visual inspections were performed as per requirement with no quality notifications related to the complaint defect.Batch 8035612 was inspected and accepted based on meeting our inspection control plan and subsequently approved for shipment.The review showed no rejected inspections or quality issues during the production of the provided lot number that could have contributed to the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.
|
|
Event Description
|
It was reported that a bd luer-lok¿ syringe sterile, single use had a plunger that was hard to pull back.There was no report of exposure, serious injury, or medical intervention.
|
|