| Model Number |
SD-240U-25 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Tissue Damage (2104)
|
| Date of Event |
10/01/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The reported device was returned to olympus for evaluation.The reported event could not be confirmed as the device passed inspection worked appropriately.A visual inspection was performed on the device and found no sharp edges, no frayed wire, or abnormalities on the physical appearance.The loop and operating wire of the device has biomaterial adhered on it which is indicative of usage but are intact.The slider was manipulated back and forth and confirmed that the loop wire extended and retracted smoothly.The device was then connected to an olympus test active cord and electrosurgical generator and it activated / functioned as intended.The cause of the reported event could not be confirmed as the concomitant equipment were not returned for testing.However, to mitigate the risk of patient injury and device damage, the instruction manual provides caution and warning which states, ¿do not insert the instrument into the endoscope unless you have a clear endoscopic field of view.If you cannot see the distal end of the insertion portion in the endoscopic field of view, do not use it.Do not activate output when the distal end of endoscope is too close to or in contact with body cavity tissue.This could burn the tissue or damage the endoscope.¿.
|
| |
|
Event or Problem Description
|
|
Olympus was informed that during the middle of a colonoscopy procedure, the device tore a patient¿s mucosa while the doctor was withdrawing the snare out of the patient¿s colon.The user facility reported there was no issue withdrawing the snare from the scope as there were no restrictions noted during insertion or withdrawal of the snare.The user facility also reported there was no treatment, intervention or additional procedure required to the patient as the user facility reported there was slight tissue tear and minor bleeding which subsided on it's own.The patient did not require a longer stay as the patient was discharged on the same day.The user facility performed a follow up with the reported patient and the patient is doing fine.No other instruments were used during the procedure.The intended procedure was completed with same equipment.The user facility reported the scope did not cause or contribute to the reported event as there was no issue noted with the scope when the reported event occurred.There was no image, partial or complete, lost during the procedure.Although they did not notice any issues during the procedure, they do not know the cause, but believe the snare contributed to the event.The snare is expected to be returned.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide additional information from the oem.The oem conducted a review of the device history records (dhr) for the concerned lot number and indicated there was no anomaly during the manufacturing of the device.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to make a correction on the procode from fas to fdi.
|
| |
|