• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION ALARIS SYRINGE MODULE SETS; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION ALARIS SYRINGE MODULE SETS; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 10798696
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No product will be returned per customer.The customer complaint could not be confirmed because the product was not returned for failure investigation.The root cause of this failure was not identified.
 
Event or Problem Description
The customer reported that they are experiencing multiple events in which the pressure sensing disc leaks during patient use.There was no patient harm.
 
Additional Manufacturer Narrative
The customer¿s report of the set leaking was confirmed.However, the leak was observed on the filter and not the pressure sensing disc.Visual inspection of the set showed that the microbore tubing had not been fully inserted into the filter outlet port.Examination under magnification showed that both sides of the tubing had insufficient solvent applied at the engagement.Functional testing confirmed leaking from the microbore tubing to the outlet port of the filter.No leaks were observed on the pressure sensing disc.Dimensional analysis was performed and the tubing measured within specification.The root cause of the customer¿s report was a manufacturing issue due to insufficient solvent being applied at the junction of the microbore tubing and the tubing not being pushed all away into the filter outlet port.
 
Event or Problem Description
The customer reported that they are experiencing multiple events in which the pressure sensing disc leaks during patient use.There was no patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARIS SYRINGE MODULE SETS
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
MDR Report Key7997198
Report Number9616066-2018-01983
Device Sequence Number289334
Product Code FPA
UDI-Device Identifier10885403234354
UDI-Public10885403234354
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
K811885
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 09/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/18/2021
Device Model Number10798696
Device Catalogue Number10798696
Device Lot Number18046218
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/02/2018
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date10/23/2018
Supplement Report FDA Received Date01/15/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
8110,8015,BD SYRINGE 60ML, TD UNK
-
-