• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHYSIO-CONTROL, INC. - 3015876 LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC. - 3015876 LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 15
Medical Device Problem Code Failure to Power Up (1476)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 10/08/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Physio-control evaluated the customer¿s device and verified the reported issue.Physio replaced the main keypad assembly.Thereafter, proper device operation was observed through functional and performance testing and the device was returned to the customer for use.
 
Event or Problem Description
The customer contacted physio-control to report that their device intermittently would not power on or off.There was no patient use associated with the reported event.
 
Additional Manufacturer Narrative
Physio-control further evaluated the removed main keypad assembly.It was observed that the left side of the button was worn and required greater force for the button press to be recognized.The right side of the button operated normally.
 
Event or Problem Description
The customer contacted physio-control to report that their device intermittently would not power on or off.There was no patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
MDR Report Key7997274
Report Number0003015876-2018-01781
Device Sequence Number18904378
Product Code MKJ
UDI-Device Identifier00883873979419
UDI-Public00883873979419
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K142430
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number15
Device Catalogue Number99577-000056
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/10/2018
Initial Date Received by Manufacturer 10/10/2018
Supplement Date Received by Manufacturer12/07/2018
Initial Report FDA Received Date10/23/2018
Supplement Report FDA Received Date12/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-