Model Number 2AF284
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/20/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product event summary: the data files and balloon catheter, 2af284 with lot number 49768 were returned and analyzed.The data files 21 applications were performed with a different balloon catheter and an electrophysiology catheter.Also, low temperature and low flow was seen in multiple applications.Visual inspection of the returned balloon catheter showed the device was intact with no apparent issues.The catheter was not easily inserted into a test laboratory sheath.The catheter passed the performance test, although the temperature was not as expected.Inflation showed a kink on the guide wire lumen under the balloon segment.Dissection showed guide wire lumen kink at 0.977 inches from the tip inside the balloons.In conclusion, the reported temperature issue was confirmed through data analysis and testing.The balloon catheter failed the returned product inspection due to a guide wire lumen kink.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that after a cryo ablation self test, the balloon catheter temperatures were not as expected.It was also reported that the balloon catheter and sheath had an "abnormally tight fit." the balloon catheter was replaced with resolve.The case was completed with cryo.No patient complications have been reported as a result of this event.The balloon catheter subsequently tested out of specification per the manufacturer's investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCTIC FRONT ADVANCE CARDIAC CRYOABLATION CATHETER
Common Device Name
PERCUTANEOUS CATHETER INTENDED FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
MDR Report Key7997321
Report Number3002648230-2018-00750
Device Sequence Number288479
Product Code OAE
UDI-Device Identifier00643169753693
UDI-Public00643169753693
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P100010/S015
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/01/2019
Device Model Number2AF284
Device Catalogue Number2AF284
Device Lot Number49768
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/31/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/28/2018
Initial Report FDA Received Date10/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-