Manufacturing review: a lot review was not performed sine no lot number was provided by the complainant.Investigation summary: one partial segment of peg safety 20 fr pull was returned for evaluation.Functional, visual, tactile and microscopic evaluations were performed.The investigation is confirmed for fluid leak, as the product had an s-shaped, circumferential crack was observed approximately 1.2 cm from one of the ends of the catheter.Also, the port was infused with a 12ml syringe and was patent to infusion and aspiration.When infused against resistance, a leak was observed at the crack.The definitive root cause could not be determined based upon available information.It is unknown if procedural issues contributed to the reported event.Labeling review: a review of product labeling documents (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) showed that the product labeling is adequate.
|