| Model Number |
496U |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
Partial thickness (Second Degree) Burn (2694)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Our local sales rep was informed that a patient received a burn to the leg that blistered.We have not been able to obtain any further details of this event.We suspect that the most likely cause of the burn injury was that the illuminated ureter probe connector, which is connected to a light cable, came into contact with the patient.Our ifu warns against attaching light cables to the patient or drape.The illuminated ureter probe was not returned for evaluation.
|
| |
|
Event or Problem Description
|
|
Allegedly, during a procedure, the patient's leg was burned by the proximal end of the ureter probe.
|
| |
|