The device was received and a gross investigation was performed.The returned device was received in an explant kit and was received disassembled in the provided plastic jar, along with the holder, the cap of the original internal container and the original plastic retainer.Following the gross investigation a visual inspection was performed.The sewing cuff was blood stained and visibly damaged with signs of cutting and residues of stitching.The orifice was broken in 4 fragments.In addition, one of the disassembled leaflets was intact meanwhile the other one was broken; only 2 fragments of the broken leaflet were returned.After the decontamination and cleaning the returned valve and its components were inspected by means of stereoscopic microscope to document their status and morphology.No additional macroscopic elements of damaging were found during this investigation step.Then, in order to better analyze the morphology of the fragments fracture surfaces, a scanning electron microscope analysis was performed.Based on the performed investigation, no material defects, voids, porosity, or inclusions were observed in the fracture surfaces that could suggest the presence of pre-existing defects.In addition to the above analysis the dhr of the carbomedics reduced valve sn (b)(4).Was retrieved and reviewed, confirming that the valve in object satisfied all material, dimensional and performance requirements of carbomedics reduced r5-021 at the time of manufacture and release.In addition, considering the type of damaging detected and the fracture locations, it is reasonably possible one of the following events occurred.The leaflet breakage may have been caused by a mishandling during the implanting / rotation procedure, likely due to the use of an incorrect rotation tool.The orifice breakage may have been caused using surgical instruments during the normal explanting procedure.Focusing the investigation on the leaflet breakage, the analysis conducted on the fracture surfaces leads to the conclusion that a load, exceeding the ultimate strength of the pyrolytic carbon, was inadvertently applied to the leaflet during the handling of the device for implantation.The above-mentioned maneuver would have caused a local over loading and bending moment of the component, finally resulting in its breakage.Given this analysis the root cause is reasonably not related to any factors intrinsic to the device in question and are likely attributable to procedural factors.Fields changed: date of report, pma/510k, if follow-up, what type,and device evaluated by mfr.
|