MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION REFURBISHED ENDOSCOPE 5.5M; LAPAROSCOPE, GENERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION REFURBISHED ENDOSCOPE 5.5M; LAPAROSCOPE, GENERAL Back to Search Results
Model Number MCENDO550R
Medical Device Problem Code Optical Problem (3001)
Health Effect - Clinical Code Blood Loss (2597)
Date of Event 09/28/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Terumo has received the device for evaluation; however, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
 
Event or Problem Description
The user facility reported to terumo cardiovascular that during vein harvesting procedure, the endoscope was cloudy.It is still unknown if the event resulted in a delay in the procedure nor if it caused or contributed to an injury to the patient.There was a blood loss but the amount is yet to be determine.Product was changed out.Procedure was completed successfully.
 
Event or Problem Description
There was a blood loss around 10 cc.There was no delay in the procedure.
 
Additional Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on oct 23, 2018.(b)(4).All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Additional Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations.(b)(4).A visual observation of the internal components using a monocular was conducted to check the performance of the endoscope.It was discovered that there was a crack on one of the internal lenses which resulted in a cloudy visual field.The most likely root cause for this failure is, the product was either dropped or inadvertently struck by an object or excessive force was exerted on the endoscope shaft; this could deflect the shaft and lead to the alteration of the internal lens system.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REFURBISHED ENDOSCOPE 5.5M
Common Device Name
LAPAROSCOPE, GENERAL
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
MDR Report Key7997965
Report Number1124841-2018-00266
Device Sequence Number11145671
Product Code GCJ
UDI-Device Identifier00699753023222
UDI-Public(01)00699753023222
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
K160206
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMCENDO550R
Device Catalogue NumberN/A
Device Lot Number742407
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 10/01/2018
Supplement Date Received by Manufacturer10/25/2018
11/14/2018
Initial Report FDA Received Date10/23/2018
Supplement Report FDA Received Date11/12/2018
11/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-