| Model Number |
MMT-751NAP |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
09/26/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Customer reported via phone call that they experienced high blood glucose.The customer blood glucose level was 423 mg/dl at the time of incident.The customer was treated with manual injection.The customer was assisted with troubleshooting and declined for high blood glucose.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed rewind test, basic occlusion test, occlusion test, prime test, excessive no delivery test and displacement test.No unexpected no delivery alarm noted during testing.(b)(4).
|
| |
|