• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751NAP; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751NAP; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-751NAP
Medical Device Problem Code No Apparent Adverse Event (3189)
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 09/26/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
Customer reported via phone call that they experienced high blood glucose.The customer blood glucose level was 423 mg/dl at the time of incident.The customer was treated with manual injection.The customer was assisted with troubleshooting and declined for high blood glucose.The insulin pump will be returned for analysis.
 
Additional Manufacturer Narrative
Device passed rewind test, basic occlusion test, occlusion test, prime test, excessive no delivery test and displacement test.No unexpected no delivery alarm noted during testing.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-751NAP
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7999015
Report Number2032227-2018-48827
Device Sequence Number1386901
Product Code OZO
UDI-Device Identifier00643169736092
UDI-Public(01)00643169736092
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberMMT-751NAP
Device Catalogue NumberMMT-751NAP
Device Lot NumberA6751NAPJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/19/2019
Initial Date Received by Manufacturer 10/10/2018
Supplement Date Received by Manufacturer06/04/2019
Initial Report FDA Received Date10/24/2018
Supplement Report FDA Received Date06/09/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
FRN RESERVOIR MMT-332AUNO INF SET
Outcome Attributed to Adverse Event Other;
Patient Age38 YR
Patient Weight91
-
-