| Catalog Number |
950502800 |
| Medical Device Problem Codes |
Peeled/Delaminated (1454); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/17/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Examination of the returned device found the instrument brackets have delaminated.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
The instruments were reported for unknown reason.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).Investigation summary: examination of the returned device found the instrument brackets have delaminated.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary : examination of the returned device found the instrument brackets have delaminated.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot : null.Device history batch : null.Device history review : null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|