| Model Number |
50610 |
| Medical Device Problem Code |
Nonstandard Device (1420)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not received.
|
| |
|
Event or Problem Description
|
|
It was reported that the coude tip of the catheter was curved more than usual and the patient was therefore unable to insert and use the catheter.No medical intervention reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "device description the magic3 go¿ intermittent urinary catheter is a silicone tube inserted into the urethra for the purpose of draining the bladder of urine.Not made with natural rubber latex does not contain dehp indications for use the magic3 go¿ intermittent urinary catheter is intended for urological use only.It is intended for use by adult female patients for bladder management including urine drainage, collection, and measurement.The device is passed to the urinary bladder via the urethra.Contraindications none known warnings this is a single use device.Do not resterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient.Precautions inspect the catheter before use.Do not use the product if the device or packaging is damaged.Adverse reactions urinary tract infection bleeding from the urethra irritation of the urethra".
|
| |
|
Event or Problem Description
|
|
It was reported that the coude tip of the catheter was curved more than usual and the patient was therefore unable to insert and use the catheter.No medical intervention reported.
|
| |
|
Search Alerts/Recalls
|