| Model Number |
60728753 |
| Medical Device Problem Codes |
Contamination /Decontamination Problem (2895); Device Contaminated During Manufacture or Shipping (2969)
|
| Health Effect - Clinical Code |
Not Applicable (3189)
|
| Date of Event |
09/30/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Bayer product analysis reviewed the photographs provided by the customer and confirmed the presence of a foreign material in the syringe barrel.In addition, the suspect product returned for evaluation.The foreign material present in the syringe barrel is white in color and appears to be plastic-like material.Our investigation determined that the particle noted in the syringe was likely introduced during the component manufacturing or assembly process.A 12 month review of complaint history determined a frequency of (b)(4).
|
| |
|
Event or Problem Description
|
|
The site reported the following: a nurse reported that a white foreign particle was found inside one of the syringes in a sds-ctp-spk kit.The kit was not used for a patient and was removed from the area to be returned for investigation.
|
| |
|