| Model Number |
TRUEMETRIX |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Codes |
Hyperglycemia (1905); Weakness (2145); Lethargy (2560)
|
| Date of Event |
09/27/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Returned meter and returned test strips evaluated with no defects found.Most likely underlying root cause: mlc-18- user has high glucose value.(b)(4).Manufacturer contacted customer on (b)(6) 2018 in a follow-up call to ensure the customer's condition since the initial call; husband states that he had to take her to the hospital.He states that when they got to the hospital and her blood glucose results was 720mg/dl.Customer stiil at the hospital.On the call back from (b)(6) 2018;customers husband states that wife is being given different medications and insulin her regiment has changed.Customers husband stated that customers blood test was 326 mg/dl earlier today.On the call back from (b)(6) 2018; customer was discharged from hospital on (b)(6) 2018.Customers husband states that she has a physical therapist and a nurse that will be coming for in home care 3 days per week.Husband states that they are still trying to get wife's blood glucose regulated.They received the replacement product.Customer is comfortable with the glucose readings from the new replacement products.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for hi display accompanied by symptoms.Husband is calling on behalf of the customer.The expected fasting blood glucose test result range is undisclosed.Customer did report symptoms of feeling lethargic and weak.Medical attention is reported as a result of the actual blood glucose results and reported symptoms.The product is stored by customer according to specification in the dining room.During the call on (b)(6) 2018, a back to back blood test was performed fasting by the customer and produced test results of hi display using true metrix meter.The test strip lot manufacturer's expiration date is 5/16/2020 and open vial date is (b)(6) 2018.The meter memory was reviewed for previous test result history: (date/time not stored correctly): result 1:hi date: (b)(6) 2018 time:7:14 pm fasting; result 2:561mg/dl date: (b)(6) 2018 time:4:08 pm fasting; result 3:308mg/dl date: (b)(6) 2018 time:10:36 pm fasting; result 4:350mg/dl date: (b)(6) 2018 time:9:23 pm fasting; result 5:288mg/dl date: (b)(6) 2018 time:9:42 pm fasting.
|
| |
|
Additional Manufacturer Narrative
|
|
(manufacturer narrative = t, corrected data = f) sections with additional information as of 29-apr-2020: h10: retention testing was performed using test strips from the same lot.Retention strip lot tested and passed.Internal report # (b)(4).Returned meter and returned test strips evaluated with no defects found.Most likely underlying root cause: mlc-18- user has high glucose value.Test strip udi: (b)(4).Note: manufacturer contacted customer on (b)(6) 2018 in a follow-up call to ensure the customer's condition since the initial call; husband states that he had to take her to the hospital.He states that when they got to the hospital and her blood glucose results was 720mg/dl.Customer stiil at the hospital.On the call back from (b)(6) 2018;customers husband states that wife is being given different medications and insulin her regiment has changed.Customers husband stated that customers blood test was 326 mg/dl earlier today.On the call back from (b)(6) 2018; customer was discharged from hospital on (b)(6) 2018.Customers husband states that she has a physical therapist and a nurse that will be coming for in home care 3 days per week.Husband states that they are still trying to get wife's blood glucose regulated.They received the replacement product.Customer is comfortable with the glucose readings from the new replacement products.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for hi display accompanied by symptoms.Husband is calling on behalf of the customer.The expected fasting blood glucose test result range is undisclosed.Customer did report symptoms of feeling lethargic and weak.Medical attention is reported as a result of the actual blood glucose results and reported symptoms.The product is stored by customer according to specification in the dining room.During the call on (b)(6) 2018, a back to back blood test was performed fasting by the customer and produced test results of hi display using true metrix meter.The test strip lot manufacturer's expiration date is 5/16/2020 and open vial date is (b)(6) 2018.The meter memory was reviewed for previous test result history: (date/time not stored correctly).Result 1:hi date: (b)(6) 2018 time:7:14 pm fasting, result 2:561mg/dl date:(b)(6) 2018 time:4:08 pm fasting, result 3:308mg/dl date:(b)(6) 2018 time:10:36 pm fasting, result 4:350mg/dl date:(b)(6) 2018 time:9:23 pm fasting, result 5:288mg/dl date:(b)(6) 2018 time:9:42 pm fasting.
|
| |
|