| Model Number |
179712760 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Unspecified Infection (1930); No Code Available (3191)
|
| Date of Event |
10/10/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # = > (b)(4).Surgical intervention in the form of a revision surgery.A complaint investigation will be performed.The complaint product is not available for the investigation.A supplemental report is not anticipated unless the results of the complaint investigation identify a corrective action or additional relevant information.Should the product become available, a physical evaluation will be conducted and a supplemental report filed with the results.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the screw was implanted in the ilium in the surgery on (b)(6) 2016.The hospital noticed that the screw had been broken in the patient¿s body before the removal surgery on (b)(6) 2018, which was performed because of infection.While the screw head was removed, the broken shaft was left in the body as planned preoperatively.There was no surgical delay.It is possible that the shaft come out from the ilium in the future.The patient is female who has parkinson¿s disease.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Visual examination at the macroscopic level found that the polyaxial screw shank had fractured.The other half of the broken screw shank was not returned for evaluation.A fracture analysis was performed.The fracture analysis exhibits two surface morphologies, a smooth region and a grainy/rough region with progression lines which are indicative of fatigue failure.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.All complaint trends will be evaluated as a part of the depuy spine monthly complaint review meeting.A definitive root cause for the polyaxial screw shank fracturing cannot be positively determined.However, the fracture analysis report exhibits two surface morphologies, a smooth region and a grainy/rough region with progression lines which are indicative of fatigue failure.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer and no systemic trends were found, this complaint file will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|