(b)(4).Physical manufacturer name: codman and shurtleff inc., dba depuy synthes products, inc.[conclusion]: the healthcare professional reported that during the coil embolization procedure of an aneurysm at the splenic artery, the deltafill 18 24mm x 60cm coil (dlf182460 / s10238) was used but the deltafill 18 coil did not make the adequate shape as the physician had expected it.When the physician attempted to resheath it, the sheath introducer became damaged.The deltafill 18 coil was replaced with another deltafill 18 of the same size and the procedure was successfully completed without further issues or delay.It was confirmed that there was no difficulty with positioning the deltafill 18 coil at the target site, no resistance at any time during the advancement of the coil through the microcatheter, there was no damage to the device inner packaging or to the actual coil prior to use, the coil delivery system was prepped without any issue, and the physician did not apply excessive force during the unsheathing and resheathing of the device.There was no report of patient injury or complication; the complaint product was new and was stored per labeling instructions.The procedure was conducted in accordance with the instruction for use (ifu) with constant flush maintained at all time.There was no visible defect or damage noted on the product prior to the event.There was no unintended detachment observed in the vessel or in the microcatheter.The product was discarded and thus, not available to be returned for evaluation.Based on complaint information, the device was not available to be returned for analysis.A review of manufacturing documentation associated with this lot (s10238) presented no issues during the manufacturing or inspection processes related to the reported complaint.All product rejected during manufacturing was identified as scrap and properly accounted for.Product analysis cannot be conducted as the product was not returned for analysis.No determination of causes and possible contributing factors could be made.As such, the investigation will be closed.With the limited information available and without the product available for analysis, the reported customer complaint could not be confirmed.Based on the device history record review, there is no indication that the event is related to the device manufacturing process.The exact cause of the event could not be conclusively determined; however, it is possible that circumstances of the procedure and/or device manipulation/interaction may have contributed to the reported failure.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective/preventive action may be triggered at a later time.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
|
|
The healthcare professional reported that during the coil embolization procedure of an aneurysm at the splenic artery, the deltafill 18 24mm x 60cm coil (dlf182460 / s10238) was used but the deltafill 18 coil did not make the adequate shape as the physician had expected it.When the physician attempted to resheath it, the sheath introducer became damaged.The deltafill 18 coil was replaced with another deltafill 18 of the same size and the procedure was successfully completed without further issues or delay.It was confirmed that there was no difficulty with positioning the deltafill 18 coil at the target site, no resistance at any time during the advancement of the coil through the microcatheter, there was no damage to the device inner packaging or to the actual coil prior to use, the coil delivery system was prepped without any issue, and the physician did not apply excessive force during the unsheathing and resheathing of the device.There was no report of patient injury or complication; the complaint product was new and was stored per labeling instructions.The procedure was conducted in accordance with the instruction for use (ifu) with constant flush maintained at all time.There was no visible defect or damage noted on the product prior to the event.There was no unintended detachment observed in the vessel or in the microcatheter.The product was discarded and thus, not available to be returned for evaluation.
|