| Catalog Number |
FFS002-01 |
| Medical Device Problem Codes |
Fracture (1260); Insufficient Heating (1287)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/17/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an ablation procedure, there were no noticeable defects there were detected during connection.However, during ablation, there was no heating and upon inspection, the laser had an opening on the side of the black cable.There were no patient complications reported in relation to this event.
|
| |
|
Event or Problem Description
|
|
It was reported that during an ablation procedure, there was no heating and upon inspection, the laser had an opening on the side of the black cable.There were no patient complications reported in relation to this event.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was inspected and the laser fiber break was midway between the connector and clampshell.The laser fiber insulation was melted and the laser fiber was completely separated; no other anomalies were observed.
|
| |
|