(b)(4).Batch # unk.The lot/batch was not provided; therefore, the manufacturing records could not be reviewed.Additional information obtained: describe the pad and patient position in the surgery procedure? supine, for cabg procedure in relation to the patient, what was the anatomical position of the rigid corner of the mega soft pad? covering the shoulders beneath the patient.What was the alignment of the pad and cord relative to the surgical site? (e.G.Was the corded side of the pad facing away from the surgical site?) since it was dual cord megasoft both the cords were connected to separate electrocautery generator on either side of the patient towards the head end.Was there anything between the pad and the patient? yes one normal draping linen between patient and the pad.Was there any patient positioning tools or aides used? if so, was the pad place above or below? yes a bandaged cushioning pad placed beneath the shoulders to lift the chest which is in between linen and patient body.How was the burn identified? the burn was identified second day of the operation as noticed by the attending nurse how was the burn treated? applied bactigras's ointment is the customer alleging that this burn is potentially associated with the use of the mega soft pad? if so, why? yes.As this incidence happened within few days after megasoft was installed in or what is the current patient status? patient is to come back for follow ups after being discharged from the hospital.
|