Catalog Number 303467
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/02/2018
Type of Reportable Event Malfunction
Event or Problem Description
(b)(6) 2018.The customer returned kit from lot 3006837 (pa) was tested in parallel with an immucor-wks retention kit from lot 3006837 (pa) on (b)(6) 2018.The customer returned kit and the retention kit from lot 3006837 (pa) gave negative control od average values in the iib/iiia rows that did not meet ifu validity criteria.The negative serum control is run in duplicate with each test run.The negative serum control should give an average od value of < or equal to 0.160 od in the iib/iiia rows.The customer complaint ((b)(4)) was reproduced.The customer was performing validation and study testing using the kit.Corrective action is being taken through fa-wks-18-008, to notify customers of the affected pakauto lot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PAKAUTO
Common Device Name
PAKAUTO
Manufacturer (Section D)
IMMUCOR GTI DIAGNOSTICS, INC.
20925 crossroads circle
waukesha WI 53186
MDR Report Key8002478
Report Number2183608-2018-00031
Device Sequence Number1531269
Product Code MYP
UDI-Device Identifier10888234500247
UDI-Public10888234500247
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
BK960059
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Administrator/Supervisor
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 09/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date12/15/2019
Device Catalogue Number303467
Device Lot Number3006837
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/27/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/20/2018
Initial Report FDA Received Date10/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-