| Catalog Number |
303467 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/02/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
(b)(6) 2018.The customer returned kit from lot 3006837 (pa) was tested in parallel with an immucor-wks retention kit from lot 3006837 (pa) on (b)(6) 2018.The customer returned kit and the retention kit from lot 3006837 (pa) gave negative control od average values in the iib/iiia rows that did not meet ifu validity criteria.The negative serum control is run in duplicate with each test run.The negative serum control should give an average od value of < or equal to 0.160 od in the iib/iiia rows.The customer complaint ((b)(4)) was reproduced.The customer was performing validation and study testing using the kit.Corrective action is being taken through fa-wks-18-008, to notify customers of the affected pakauto lot.
|
| |
|