| Model Number |
ZXT300 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown, not provided.Explant date: if explanted, give date: unknown, information not provided.(b)(4).Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to johnson and johnson surgical vision has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the tecnis symfony toric multifocal intraocular lens (model zxt300, +20.0 diopter) was explanted from the patient¿s left eye because of patient experiencing double vision.There was no patient injury.Reportedly a model zct450 and +21.0 diopter was used as the replacement lens for the patient.No further information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the product testing could not be performed as the product was not returned.The reported complaint cannot be confirmed.Manufacturing record review: the manufacturing process record was evaluated and no deviation was found during process related to the complaint issue reported.There was no discrepancy found during the mrr (manufacturing record review).The product was manufactured and released according to specifications.A search on complaints revealed that no similar complaints were received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-010215.
|
| |
|