| Model Number |
ZXT150 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
09/11/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If implanted; give date: n/a (not applicable).No patient involvement.If explanted; give date: n/a (not applicable).No patient involvement.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a hair was found under a zxt150 intraocular lens (iol) during product handling and prior to insertion.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device available for evaluation: yes.Returned to manufacturer on 11/16/2018.Device returned to manufacturer: yes.Device evaluation: only the lens box was received at the manufacturing site on 11/16/2018.The product was not returned; therefore, the complaint issue reported was not verified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.Labeling review: the directions for use (dfu) were reviewed.The dfu adequately provides instructions, precautions, and warnings for the proper use and handling of the product.Historical data analysis: a search of complaints revealed no additional investigation requests have been received for this production order number.Conclusion: as a result of the investigation there is no indication of a quality product deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
|
| |
|