JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS CL; MONOFOCAL IOLS Back to Search Results
Model Number Z9002
Medical Device Problem Code Separation Problem (4043)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/27/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If implanted; explanted; give date: not applicable as the lens was inserted and removed.The intraocular lens (iol) is not returning for evaluation as it was discarded by the customer; therefore a failure analysis of the complaint device cannot be completed.A review of the device history record, labeling, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Attempts have been made to obtain missing information; however, to date, no response has been received.
 
Event or Problem Description
It was reported that during the implantation of the z9002 intraocular lens (iol) into the patient's left eye, the trailing haptic was noticed to be detached.Additional information was received and it was learnt that the iol was fully inserted into the eye, removed and replaced with another iol during the same procedure.The incision was not enlarged, no sutures were used and no secondary procedure was required.Reportedly, the patient tolerated the procedure well, no post-op complications.The lens was discarded by the customer.No further information was provided.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS CL
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA 92705
MDR Report Key8004007
Report Number2648035-2018-01391
Device Sequence Number13729395
Product Code HQL
UDI-Device Identifier05050474530744
UDI-Public(01)05050474530744(17)211024
Combination Product (Y/N)Y
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P880081
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/04/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/24/2021
Device Model NumberZ9002
Device Catalogue NumberZ900200225
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 09/27/2018
Supplement Date Received by Manufacturer10/25/2020
Initial Report FDA Received Date10/25/2018
Supplement Report FDA Received Date11/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-