• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VASORUM LTD CELT ACD VASCULAR CLOSURE DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VASORUM LTD CELT ACD VASCULAR CLOSURE DEVICE Back to Search Results
Model Number KCLT-6F
Medical Device Problem Code Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 09/21/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
During the process of this complaint, attempts were made to obtain complete event, patient and device information.
 
Event or Problem Description
Device was deployed according to the proper steps, but wasn't pulled all the way to the arteriotomy before turning step 2.It was then released in the vessel.Doctor re-accessed in the same location and placed a 6 x/ 40mm zilver stent to capture the device against the wall of the sfa.
 
Additional Manufacturer Narrative
During the process of this complaint, attempts were made to obtain complete event, patient and device information.Vasorum ltd is submitting this report on the behalf of (b)(4), in accordance with the conditions outlined with the mdr exemption e2017048.Vasorum ltd, trinitas house, 2012 orchard avenue, citywest business park, d24 wpw6, dublin - ireland.(b)(4).
 
Event or Problem Description
Device was deployed according to the proper steps, but wasn't pulled all the way to the arteriotomy before turning step 2.It was then released in the vessel.Doctor re-accessed in the same location and placed a 6 x/ 40mm zilver stent to capture the device against the wall of the sfa.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELT ACD VASCULAR CLOSURE DEVICE
Common Device Name
VASCULAR CLOSURE DEVICE
Manufacturer (Section D)
VASORUM LTD
2012 orchard avenue
citywest business park
dublin, leinster D24 W PW6
EI  D24 WPW6
MDR Report Key8004263
Report Number3009984513-2018-00002
Device Sequence Number19179198
Product Code MGB
UDI-Device Identifier05391530280082
UDI-Public(01)05391530280082(17)210628(10)654934
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P150006
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Patient Monitoring
Type of Report Initial,Followup
Report Date (Section B) 04/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/28/2021
Device Model NumberKCLT-6F
Device Lot Number654934
Was Device Available for Evaluation? No
Distributor Facility Aware Date10/05/2018
Device Age3 MO
Initial Date Received by Manufacturer 10/05/2018
Supplement Date Received by Manufacturer10/05/2018
Initial Report FDA Received Date10/25/2018
Supplement Report FDA Received Date04/04/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
Patient Weight84
-
-