| Catalog Number |
FFS222-01 |
| Medical Device Problem Codes |
Fracture (1260); Insufficient Heating (1287)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/24/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation: confirmation that the pilot light was not working by plugging the probe into the test system.The pilot light did not pass through the probe indicating a broken fiber.
|
| |
|
Event or Problem Description
|
|
It was reported that during an ablation procedure, the user pulled the probe out of the packaging and set to the correct depth.The user then plugged in the e2k connector fully to check the pilot light and it was noted that there was no pilot light present.The user then switched probes out.There were no patient complications reported in relation to this event.
|
| |
|