• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MONTERIS MEDICAL FULLFIRE SELECT DIFFUSING TIP LASER PROBE; FULLFIRE¿ SELECT DIFFUSING TIP LASER PROBE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MONTERIS MEDICAL FULLFIRE SELECT DIFFUSING TIP LASER PROBE; FULLFIRE¿ SELECT DIFFUSING TIP LASER PROBE Back to Search Results
Catalog Number FFS122-01
Medical Device Problem Codes Fracture (1260); Insufficient Heating (1287)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/30/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluation: the probe was plugged into the test system and the pilot light did not light up; fiber confirmed broken.
 
Event or Problem Description
It was reported that a probe was opened during an ablation procedure and contained a broken fiber.The probe was inserted and fired, however no heating was observed.The probe was then removed and no pilot light was visible at the probe tip.The probe was replaced and the case proceeded successfully.There were no patient complications reported in relation to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FULLFIRE SELECT DIFFUSING TIP LASER PROBE
Common Device Name
FULLFIRE¿ SELECT DIFFUSING TIP LASER PROBE
Manufacturer (Section D)
MONTERIS MEDICAL
14755 27th ave n
suite c
plymouth MN 55447
MDR Report Key8005201
Report Number3009970070-2018-00082
Device Sequence Number19130256
Product Code GEX
UDI-Device Identifier00816589020700
UDI-Public00816589020700
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K081509
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/05/2018
Device Catalogue NumberFFS122-01
Device Lot NumberC8845-09
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/24/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/30/2017
Initial Report FDA Received Date10/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-