• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MONTERIS MEDICAL NEUROBLATE® SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MONTERIS MEDICAL NEUROBLATE® SYSTEM Back to Search Results
Model Number 20360
Medical Device Problem Code Computer Software Problem (1112)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/17/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an ablation procedure, blue pixel artifact were witnessed.The user was ablating using firing power of 128% and the user noticed blue pixels about 1cm from the probe.The user stopped delivering laser and the blue pixels slowly diminished.The user went back on at 100% and the blue pixels came back.The user noted the blue pixel artifact and proceeded through the entire case with the blue pixel artifact.There were no patient complications reported in relation to this event.
 
Additional Manufacturer Narrative
Additional information: device evaluation: the blue pixels in this case were caused by changes in the tissue that appear in the magnitude data as voids that move through the image.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUROBLATE® SYSTEM
Common Device Name
NEUROBLATE® SYSTEM
Manufacturer (Section D)
MONTERIS MEDICAL
14755 27th ave n
suite c
plymouth MN 55447
MDR Report Key8005219
Report Number3009970070-2018-00081
Device Sequence Number13750651
Product Code GEX
UDI-Device Identifier00816589020656
UDI-Public00816589020656
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K173305
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number20360
Device Catalogue NumberNB102
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 10/17/2018
Supplement Date Received by Manufacturer11/15/2018
Initial Report FDA Received Date10/25/2018
Supplement Report FDA Received Date12/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-