Reportable based on device analysis completed on 04oct2018.This 3.00mm x 8mm emerge balloon catheter was received with no reported issues.However, returned device analysis revealed separated hypotube.Device evaluated by manufacturer: the returned product consisted of an emerge balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed multiple kinks along the hypotube.The hypotube is separated 79.6cm from the hub.Microscopic examination revealed that the tip is damaged and the balloon is loose.There is blood present in the hub, inflation lumen, and balloon.The broken shaft was connected to a tuohy adapter and stopcock then inflated using a balloon inflation device to check for possible damages to the balloon.The balloon was inflated to rbp of 14atm and could hold pressure for 5 minutes before it was deflated.Inspection of the remainder of the device presented no other damage or irregularities.
|