Model Number 7444
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/28/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation, however, images were provided.A manufacturing record review was completed and zero nonconformances were found therefore supporting the device met material, assembly and performance specifications.Based on the information provided stating the wire knuckled in a stent and became knotted, causing the coils to unravel/separate from the core wire, the most likely root cause for the damage is operational context (device interaction with stent).
 
Event or Problem Description
While physician was using the bandit guidewire, it knuckled in a stent.The wire became knotted when he knuckled it, and the coils uncoiled/separated from the core wire.The device remained intact and no pieces of the wire were left in the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BANDIT GUIDEWIRE
Common Device Name
CATHETER GUIDEWIRE
Manufacturer (Section D)
VASCULAR SOLUTIONS, LLC
6464 sycamore court north
minneapolis MN 55369
MDR Report Key8006248
Report Number2134812-2018-00077
Device Sequence Number1150987
Product Code DQX
UDI-Device IdentifierM20674440
UDI-PublicM20674440
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K181647
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/23/2020
Device Model Number7444
Device Lot Number632529
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/28/2018
Initial Report FDA Received Date10/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-