Model Number S403
Medical Device Problem Codes Over-Sensing (1438); Pacing Problem (1439)
Health Effect - Clinical Code Syncope (1610)
Date of Event 09/28/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event or Problem Description
It was reported that this pacemaker was oversensing atrial activity on the ventricular channel resulting in pacing inhibition for a dependent patient.As a result of the pacing inhibition the patient became syncopal.The blanking period after atrial activity was extended but did not resolve the issue.The ventricular sensitivity was modified to maximum sensitivity which improved the oversensing but did not eliminate it.The physician reported that there were no problems with lead placement based on x-ray imaging.The patient was monitored overnight in the hospital and discharged the following day.No surgical intervention was performed and the device remains in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALTRUA 40
Common Device Name
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
MDR Report Key8006315
Report Number2124215-2018-60485
Device Sequence Number203553
Product Code LWP
UDI-Device Identifier00802526484445
UDI-Public00802526484445
Combination Product (Y/N)N
Initial Reporter CountryMX
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/23/2013
Device Model NumberS403
Device Catalogue NumberS403
Device Lot Number588310
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/28/2018
Initial Report FDA Received Date10/25/2018
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Life Threatening; Required Intervention;
-
-