| Catalog Number |
6020-4535 |
| Medical Device Problem Codes |
Degraded (1153); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Edema (1820); Inflammation (1932); Injury (2348); Reaction (2414)
|
| Date of Event |
09/27/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history records indicates that the reported devices were manufactured and accepted into final stock with no reported discrepancies.The complaint history review indicated that there were no similar events for the reported lot.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Not returned to the manufacturer.
|
| |
|
Event or Problem Description
|
|
The metal head was worn off and ruined the stems trunnion.Update 16/october/2018: rep listed the following adverse consequences: delay in surgery, metallosis, pseudotumor.
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding disassociation involving an accolade stem was reported.The event was confirmed through x--ray and photographs of the explanted device.Method & results: product evaluation and results: no product was returned for evaluation, however photographs were provided for review.The photographs show a recently explanted stem and head.The neck/trunion of the stem is worn, there is also blackening on the stem.Clinician review: a review of the provided medical records by a clinical consultant stated the following comment: x-ray and photo confirms disassociation and trunion erosion/corrosion.Product history review: indicated all devices were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusion: a review of the provided medical records by a clinical consultant stated the following comment: x-ray and photo confirms disassociation and trunnion erosion/corrosion further information such as operative reports, patient demographics, past medical history and examination of explanted products are required to complete the investigation for determining a root cause.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Event or Problem Description
|
|
The metal head was worn off and ruined the stems trunnion.Update (b)(6) 2018: rep listed the following adverse consequences: delay in surgery, metallosis, pseudotumor.
|
| |
|
Search Alerts/Recalls
|