Catalog Number 3992208
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the r eported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient was pre-operatively diagnosed with lumbar canal stenosis and underwent minimally invasive surgery (mis) - transforaminal lumbar interbody fusion (tlif) at l2,3,4,5.Post-op, at l2/3, the spacer backed out.No patient symptoms or complication were reported as result of this event.
 
Additional Manufacturer Narrative
Image review: ap/lateral x-rays for l2-l5 plif/spinal stabilization shows an extended interbody graft at l2-3.There is solid fusion at l4-5 but a paucity of bone at l3-4/l2-3 otherwise hardware placement appears appropriate if information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSTONE PTC SPINAL SYSTEM
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
Manufacturer (Section D)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
MDR Report Key8007163
Report Number1030489-2018-01406
Device Sequence Number203568
Product Code MAX
UDI-Device Identifier00643169187559
UDI-Public00643169187559
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K172199
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/25/2024
Device Catalogue Number3992208
Device Lot NumberH5227840
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/27/2018
Supplement Date Received by Manufacturer11/27/2018
Initial Report FDA Received Date10/25/2018
Supplement Report FDA Received Date12/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-