| Catalog Number |
3992208 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the r eported event.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient was pre-operatively diagnosed with lumbar canal stenosis and underwent minimally invasive surgery (mis) - transforaminal lumbar interbody fusion (tlif) at l2,3,4,5.Post-op, at l2/3, the spacer backed out.No patient symptoms or complication were reported as result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Image review: ap/lateral x-rays for l2-l5 plif/spinal stabilization shows an extended interbody graft at l2-3.There is solid fusion at l4-5 but a paucity of bone at l3-4/l2-3 otherwise hardware placement appears appropriate if information is provided in the future, a supplemental report will be issued.
|
| |
|