BECTON, DICKINSON & CO. BD VACUTAINER® GLASS CPT MOLECULAR DIAGNOSTICS TUBE; LYMPHOCYTE SEPARATION MEDIUM Back to Search Results
Catalog Number 362753
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/04/2018
Event Type  malfunction  
Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported with the use of the bd vacutainer® glass cpt molecular diagnostics tube there was an issue with poor separation.There was no report of injury or medical intervention.
 
Event Description
It was reported with the use of the bd vacutainer® glass cpt molecular diagnostics tube there was an issue with poor separation.There was no report of injury or medical intervention.
 
Manufacturer Narrative
Investigation summary: bd had not received samples, but photos were provided by the customer facility for investigation.The photos were evaluated and the customer's indicated failure mode for poor barrier with the incident lot was observed.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Bd has initiated further investigation relating to this issue through a capa.The investigation is still on-going and improvements will be made as the potential causes of this issue are identified.Based on evaluation of the customer photos, the customer¿s indicated failure mode for poor barrier with the incident lot was observed.Further investigation has been initiated through a capa.The investigation is still on-going and improvements will be made as the potential causes are identified.A capa has been initiated to document further investigation and root cause analysis relating to this issue.The investigation is currently on-going and will be updated as the potential root cause(s) are identified.Based on an assessment of severity and frequency, it was determined that a capa is required at this time in order to determine the root cause associated with this issue and implement corrective actions.The investigation is currently on-going and further improvements will be made as the potential root cause(s) are identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® GLASS CPT MOLECULAR DIAGNOSTICS TUBE
Type of Device
LYMPHOCYTE SEPARATION MEDIUM
Manufacturer (Section D)
BECTON, DICKINSON & CO.
150 south 1st avenue
broken bow NE 68822
MDR Report Key8007288
MDR Text Key125141019
Report Number1917413-2018-03660
Device Sequence Number1
Product Code JCF
UDI-Device Identifier50382903627535
UDI-Public50382903627535
Combination Product (y/n)N
PMA/PMN Number
K891407
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Type of Report Initial,Followup
Report Date 11/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/25/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date06/30/2019
Device Catalogue Number362753
Device Lot Number8152655
Date Manufacturer Received10/05/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-