| Medical Device Problem Codes |
Image Display Error/Artifact (1304); Structural Problem (2506)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation: the pump has been returned and evaluated by product analysis on 11-oct-2018 with the following findings: during investigation, it was found that the display was dim with a pink hue.Further investigation found that the battery compartment was cracked from the top above the pad to the cover part.(b)(6).
|
| |
|
Event or Problem Description
|
|
The pump was returned for investigation.Investigation revealed the pump had a dim display with a pink hue.Investigation revealed the battery compartment was cracked from the top above the pad to the cover part.This report is made based on results of investigation completed on 11-oct-2018.
|
| |
|