BOSTON SCIENTIFIC CORPORATION ACCUMAXTM SINGLE USE HOLMIUM LASER FIBER; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number M0068404012
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/27/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device has been received for analysis; however the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation on (b)(4) 2018 that an accumax laser fiber was used during a ureteroscopy procedure performed on (b)(6) 2018.According to the complainant, during procedure, the tip of the laser fiber was noted to be broken straight from the package.Attempts to obtain additional information regarding this event have been unsuccessful to date.Should additional relevant details become available; a supplement report will be submitted.
 
Additional Manufacturer Narrative
Investigation result: one flexiva 200 tractip laser fiber was returned broken into two pieces with the tip detached.The first piece measured approximately 95.1 cm from the top of the connector to the break.The second piece measured approximately 211.2 cm from the top of break to the tip.The distal tip was not returned.The pattern on the tip face is consistent with a tip detachment, likely as a result of handling during the procedure.Visual inspection of the returned fiber noted burn marks on the fiber jacketing, likely a result of the fiber break occurring during use, resulting in a misfire.Review and analysis of all available information indicated that the root cause is manufacturing deficiency.A complaint with a cause of manufacturing deficiency indicates that the problems were traced to the manufacturing process.An investigation was completed to address this issue which identified the most probable root cause of the fiber break as handling during the manufacturing process caused by the supplier manufacturer.A review of the device history record (dhr) was performed and no deviations were found.
 
Event or Problem Description
It was reported to boston scientific corporation on october 2, 2018 that an accumax laser fiber was used during a ureteroscopy procedure performed on (b)(6) 2018.According to the complainant, during procedure, the tip of the laser fiber was noted to be broken straight from the package.Attempts to obtain additional information regarding this event have been unsuccessful to date.Should additional relevant details become available; a supplement report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCUMAXTM SINGLE USE HOLMIUM LASER FIBER
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key8010861
Report Number3005099803-2018-61075
Device Sequence Number165287
Product Code GEX
UDI-Device Identifier08714729784623
UDI-Public08714729784623
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K093691
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/06/2021
Device Model NumberM0068404012
Device Catalogue Number840-401
Device Lot Number0000006432
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/23/2018
Initial Date Received by Manufacturer 10/02/2018
Supplement Date Received by Manufacturer01/09/2019
Initial Report FDA Received Date10/26/2018
Supplement Report FDA Received Date01/28/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-