• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MDT POWERED SURGICAL SOLUTIONS LEGEND FOOTED ATTACHMENT; MOTOR, DRILL, PNEUMATIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT POWERED SURGICAL SOLUTIONS LEGEND FOOTED ATTACHMENT; MOTOR, DRILL, PNEUMATIC Back to Search Results
Model Number AF01-R
Medical Device Problem Code No Apparent Adverse Event (3189)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/11/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Report confirmed.Evaluation determined that a portion of the foot of the attachment was detached.The likely cause of failure was due to inadequate preventative maintenance.It was also noted that the internal tube bearing was corroded.The user manual contains the following warnings ¿do not use a legend attachment if any part of the attachment appears to be bent, loose, missing or damaged.Do not use excessive pressure, such as bending or prying, on attachments or dissecting tools.This may cause tool to bend or break and cause injury to patient, operator and/or operating room staff." in addition, ¿insert dissecting tool into motor collet with a slight rotational motion.A tactile and audible click is observed indicating that the dissecting tool is fully seated the maximum specified service interval is 24 months.Device has been in use for approximately 75 months with no record of factory service during this period.We will continue to track and trend this complaint type.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Repair request initiated for device with the report of not working properly.It was reported that there was no patient impact.Upon decontamination, it was determined that foot was detached.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEGEND FOOTED ATTACHMENT
Common Device Name
MOTOR, DRILL, PNEUMATIC
Manufacturer (Section D)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach street
fort worth TX 76137
MDR Report Key8011105
Report Number1625507-2018-00066
Device Sequence Number170065
Product Code HBB
UDI-Device Identifier00613994422217
UDI-Public00613994422217
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K020069
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAF01-R
Device Catalogue NumberAF01-R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/20/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/04/2018
Initial Report FDA Received Date10/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-