| Catalog Number |
100/897/080 |
| Medical Device Problem Codes |
Decrease in Pressure (1490); Device Contamination with Body Fluid (2317)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/18/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Report source: (b)(6).
|
| |
|
Event or Problem Description
|
|
Information was received that a smiths medical portex uniperc adjustable flange tracheostomy tube had problems with the cuff holding pressure while in use with a patient.Twenty days after incretion, the tracheostomy (trach) tube was replaced with another smiths medical portex trach.Following the change out, it was discovered the initial trach pilot balloon and cuff appeared to be full of "water, fluid, and/or secretions." no adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
One portex uniperc tracheostomy (trach) tube was returned for analysis in used conditions.Under visual inspection the trach tube appeared to be in good condition, no water / fluid was found in the pilot balloon and no tear in the cuff was noted.Inflation test was performed; no discrepancies found.Based on the evidence there was no fault found for the root cause as the complaint was not confirmed.
|
| |
|