| Model Number |
M00534680 |
| Medical Device Problem Code |
Unsealed Device Packaging (1444)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
10/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The exact date of this event is unknown.The provided event date was chosen as a best estimate based on the date that the manufacturer became aware of this event.(b)(6).(b)(4).Although the suspect device has been received, the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that when the device was received, it was noticed that the packaging of an advanix biliary stent was broken and the sterility of the device was compromised.The exact date of this event was not reported.There was no patient or procedure involved.
|
| |
|
Additional Manufacturer Narrative
|
|
A visual evaluation of the device found that a clear plastic side of the pouch where the label goes placed was found to be broken, also the label was broken and the product has wrinkles indicating the manipulation of the unit.Based on the product analysis, it is possible that the way in which the device was handled and manipulated during the processes of shipping/receiving and/or storage may have contributed to the failure encountered (pouch and label broken); the defect found is typically caused due to incorrect manipulation of the product while its transport, taking into consideration that the unit was returned inside its original package.Additionally, the returned product has wrinkles indicating the manipulation of the unit.Therefore,cause traced to transport/storage is selected as the most probable root cause for the complaint.A dhr review was performed and no anomalies were observed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that when the device was received, it was noticed that the packaging of an advanix biliary stent was broken and the sterility of the device was compromised.The exact date of this event was not reported.There was no patient or procedure involved.
|
| |
|