ANIMAS LLC ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Medical Device Problem Code Structural Problem (2506)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
 
Event or Problem Description
On (b)(6) 2018, the reporter contacted animas, alleging a casing/condition (cracked/damaged casing) issue; cracked battery compartment.The reporter denied damage to the battery cap.There is no indication the alleged product issue caused or contributed to an adverse event.This complaint is being reported because the alleged malfunction can result in a delay in treatment or long term cessation in delivery if the damage impacts the power circuit or cartridge compartment.
 
Additional Manufacturer Narrative
Device evaluation: the device has been returned and evaluated by product analysis on 07-jan-2019 with the following findings: during investigation, the battery compartment was cracked above and below the grip pad.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Common Device Name
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS LLC
965 chesterbrook blvd
wayne PA 19087
MDR Report Key8011274
Report Number2531779-2018-18833
Device Sequence Number13849339
Product Code OYC
UDI-Device Identifier10840406100181
UDI-Public10840406100181
Combination Product (Y/N)Y
Initial Reporter StateWY
Initial Reporter CountryUS
PMA/510(K) Number
P130007
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2019
Was the Report Sent to FDA? Yes
Device Age23 MO
Initial Date Received by Manufacturer 10/18/2018
Supplement Date Received by Manufacturer10/18/2018
Initial Report FDA Received Date10/28/2018
Supplement Report FDA Received Date01/10/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age24 YR
Patient Weight150
-
-