Model Number N/A
Device Problem Failure to Calibrate (2440)
Patient Problem No Patient Involvement (2645)
Event Date 10/04/2018
Event Type  malfunction  
Manufacturer Narrative
The production device history record (dhr) for this intra-aortic balloon pump (iabp) was not required to be reviewed per getinge standard operating procedure since the device manufacture date is greater than one year from the event date.To address the failure, the getinge fse removed and replaced the faulty video generator board.The iabp was unit reassembled and the pm was completed.The iabp unit was then cleared for clinical use.The full name of the initial reporter listed is (b)(64); which differs from that of the event site.The full name of the event site is (b)(6).
 
Event Description
It was reported that during a preventive maintenance (pm) performed by a getinge field service engineer (fse), the cardiosave intra-aortic balloon pump (iabp) failed the touchscreen calibration.There was no patient involvement; thus, no adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer Contact
1300 macarthur blvd.
mahwah, NJ 07430
MDR Report Key8011425
MDR Text Key125297579
Report Number2249723-2018-01867
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 10/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/28/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberN/A
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Date Manufacturer Received10/04/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-