Catalog Number 48511820B
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/04/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the tip of screwdriver shaft broke off into the screwdriver/screw interface.
 
Additional Manufacturer Narrative
Method: visual inspection, product history review, complaint history review, nc/capa history review, labelling review, risk assessment.Result: the reported event was confirmed upon visual inspection.Manufacturing files were reviewed and no anomalies were found.Manufacturing review revealed an instrument over 9 years old.Conclusion: the probable root cause is normal wear based on instrument age.
 
Event or Problem Description
It was reported that the tip of screwdriver shaft broke off into the screwdriver/screw interface.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REFLEX HYBRID QUICK TURN INNER SHAFT
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
STRYKER SPINE-US
2 pearl court
allendale NJ 07401
MDR Report Key8011427
Report Number0009617544-2018-00248
Device Sequence Number13795213
Product Code LXH
UDI-Device Identifier07613327277012
UDI-Public07613327277012
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number48511820B
Device Lot Number088066
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/31/2018
Initial Date Received by Manufacturer 10/04/2018
Supplement Date Received by Manufacturer12/27/2018
Initial Report FDA Received Date10/28/2018
Supplement Report FDA Received Date01/04/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-