| Model Number |
MODEL 3000 |
| Medical Device Problem Code |
Computer Software Problem (1112)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/11/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
Physician reported that the patient's device settings had changed without intentional programming by the physician.
|
| |
|
Additional Manufacturer Narrative
|
|
Unique identifier (udi) #", corrected data: follow-up report #1 inadvertently listed the wrong udi#.
|
| |
|
Event or Problem Description
|
|
Additional investigation of this issue was performed via review of log files and it was revealed that this issue was due to a partial programming.During programming of new output currents, the output current is first cleared to 0 ma and then the new output current is programmed.However, due to disrupting in this programming process, the device output currents were cleared to 0 ma and the new output currents were not successfully programmed.Per labeling, if any parameter changes were made during a patient visit, it is recommended to perform a final interrogation prior to the end of the patient visit in order to confirm the generator is set to the desired values.
|
| |
|