We have not yet received the device for evaluation.Hence, we could not conclusively determine the root cause of the defect.It is possible that the tortuous anatomy of the patient, calcified plaque if present in the patient's vessel and excessive tension on the device during the procedure could had led to this failure.As stated in ifu, to minimize the risk of balloon rupture/separation, maximum recommended inflation volume and pull force for the catheter should not be exceeded.We have conducted a lot history review and did not find any issues noted in the manufacturing or packaging process that could be related to this issue.All qc tests passed their requirements.We perform 100% inspection on each device to ensure that the ligature properly binds to the shaft and an adequate amount of glue is present in the ligature.However, it is also possible that the tension on the ligature or the glue applied underneath the balloon during the balloon tying process was inadequate in this device resulting in the unwinding of the ligature.It could not be determined if some anatomical or procedural factors may have contributed to the balloon rupture.We have requested for additional information from the hospital related to this incident.However, we have not received any response back from the hospital.We currently have a corrective and preventive action (capa) to address this issue.
|