Model Number URF-V2R
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The customer returned urf-v2r scope to olympus for evaluation.A visual and microscopic inspection was performed on the returned device and found no signs of metal exposed or protruding from the bending section cover.However, the bending section cover was noted to be broken.The scope¿s bending section cover was removed and found the skeleton tab was completely broken/detached.No sharp edges noted with the broken/detached skeleton.Based on the evaluation the cause of the broken skeleton is due to excessive stress and force.The instruction manual warns users ¿do not twist or bend the bending section with your hands.Equipment damage may result; do not insert the insertion tube with excessive force into the ureter or calix.The bending section may be damage.¿.
 
Event Description
The user facility informed olympus that the scope¿s skeleton was protruding from the bending section of scope.No further information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO VIDEOSCOPE
Type of Device
URETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
connie tubera
2400 ringwood avenue
san jose, CA 95131
4089355124
MDR Report Key8011486
MDR Text Key125460367
Report Number2951238-2018-00658
Device Sequence Number1
Product Code FGB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K072957
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 10/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-V2R
Device Catalogue NumberURF-V2R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/10/2018
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/10/2018
Initial Date FDA Received10/28/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-