Model Number ZXT150
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Capsular Bag Tear (2639); No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was indicated that the device is not returning for evaluation; therefore a failure analysis of the complaint device cannot be completed.A review of the device/lot history record, labeling, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that a zxt150 23.5 diopter lens was implanted and did not center correctly and there was an attempt to reposition the lens.Reportedly, the haptic was damaged during the original folding, which caused the patient's capsular bag to rupture.The damaged haptic detached and became stuck/adhered to the capsular bag couldn't be removed.The lens had to be cut out later.This event may be due to a use error during handling/preparation of the lens for implant.No additional information was provided.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY TORIC
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA
MDR Report Key8011494
Report Number9614546-2018-01017
Device Sequence Number1585362
Product Code POE
UDI-Device Identifier05050474605275
UDI-Public(01)05050474605275(17)230314
Combination Product (Y/N)Y
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/04/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/14/2023
Device Model NumberZXT150
Device Catalogue NumberZXT150U235
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 09/27/2018
Supplement Date Received by Manufacturer10/25/2020
Initial Report FDA Received Date10/28/2018
Supplement Report FDA Received Date11/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Date Report Sent to FDA01/01/2005
-
-