| Model Number |
PCB00 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
09/28/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Age/date of birth: not applicable.Sex/gender: not applicable.If implanted, give date: not applicable as the lens was not implanted if explanted, give date: not applicable as the lens was not implanted, therefore not explanted.Attempts have been made to obtain missing information; however, no definitive response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a pcb00 intraocular lens (iol) when removed from packaging had broken haptic.There was no patient contact.No other information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device available for evaluation ¿ yes, returned to manufacturer on 11/27/2018.Device returned to manufacturer ¿ yes.Device evaluation: a visual inspection and evaluation using magnification was performed.Residues of viscoelastic material were not observed.No assembly error and/or defect were observed in the preloaded device related to manufacturing process.The condition in which the sample returned is consistent with a product that was handled and prepared for a surgical process.As part of the visual inspection, the preloaded device was opened to address the complaint issue reported: the lens was observed at the lens storage chamber in good conditions, also the haptics.The complaint issue haptic damaged was not confirmed.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.Labeling review: the directions for use (dfu) were reviewed.The dfu adequately provides instructions, precautions, and warnings for the proper use and handling of the product.Historical data analysis: a search on complaints related to this production order (po) was conducted.The search revealed one (1) additional complaint issue received for this po.The status closed, ¿not a device problem¿, product deficiency not identified.Conclusion: as a result of the investigation there is no indication of a quality product deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-010215.
|
| |
|