JOHNSON AND JOHNSON SURGICAL VISION, INC. TECNIS ITEC PRELOADED 1-PIECE IOL; MONOFOCAL IOLS Back to Search Results
Model Number PCB00
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 09/28/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Age/date of birth: not applicable.Sex/gender: not applicable.If implanted, give date: not applicable as the lens was not implanted if explanted, give date: not applicable as the lens was not implanted, therefore not explanted.Attempts have been made to obtain missing information; however, no definitive response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that a pcb00 intraocular lens (iol) when removed from packaging had broken haptic.There was no patient contact.No other information was provided.
 
Additional Manufacturer Narrative
Device available for evaluation ¿ yes, returned to manufacturer on 11/27/2018.Device returned to manufacturer ¿ yes.Device evaluation: a visual inspection and evaluation using magnification was performed.Residues of viscoelastic material were not observed.No assembly error and/or defect were observed in the preloaded device related to manufacturing process.The condition in which the sample returned is consistent with a product that was handled and prepared for a surgical process.As part of the visual inspection, the preloaded device was opened to address the complaint issue reported: the lens was observed at the lens storage chamber in good conditions, also the haptics.The complaint issue haptic damaged was not confirmed.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.Labeling review: the directions for use (dfu) were reviewed.The dfu adequately provides instructions, precautions, and warnings for the proper use and handling of the product.Historical data analysis: a search on complaints related to this production order (po) was conducted.The search revealed one (1) additional complaint issue received for this po.The status closed, ¿not a device problem¿, product deficiency not identified.Conclusion: as a result of the investigation there is no indication of a quality product deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS ITEC PRELOADED 1-PIECE IOL
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON AND JOHNSON SURGICAL VISION, INC.
santa ana CA 92705
MDR Report Key8011613
Report Number2648035-2018-01400
Device Sequence Number1379874
Product Code HQL
UDI-Device Identifier05050474558427
UDI-Public(01)05050474558427(17)210612
Combination Product (Y/N)Y
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/04/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/12/2021
Device Model NumberPCB00
Device Catalogue NumberPCB0000280
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/27/2018
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 09/28/2018
Supplement Date Received by Manufacturer11/30/2018
10/25/2020
Initial Report FDA Received Date10/28/2018
Supplement Report FDA Received Date12/29/2018
11/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-