• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT HI-TORQUE VERSACORE MODIFIED J GUIDE WIRE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT HI-TORQUE VERSACORE MODIFIED J GUIDE WIRE SYSTEM Back to Search Results
Catalog Number 1012068-09
Medical Device Problem Codes Difficult to Remove (1528); Material Separation (1562); Stretched (1601)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/09/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for analysis.The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that interaction with the anatomy and/or other devices resulted in the reported difficulty to remove.Manipulation of the device resulted in the reported stretched coils and ultimately resulted in the reported detachment.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot specific quality issue.There is no indication of a product quality issue with respect to manufacture, design or labeling of the device.The additional versacore device referenced is being filed under separate medwatch report.
 
Event or Problem Description
It was reported that two versacore guide wires were used in the same procedure.Resistance was felt with the anatomy during the removal of both versacore guide wires, and the guide wire tips felt like they were unraveling.Both guide wires were removed from the anatomy as a single unit, and the wire separated into two pieces outside of the patient anatomy.No intervention was required to retrieve the separated portion.Another unspecified guide wire was used to complete the procedure.There was no adverse patient effect and no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HI-TORQUE VERSACORE MODIFIED J GUIDE WIRE SYSTEM
Common Device Name
GUIDE WIRE
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key8011701
Report Number2024168-2018-08226
Device Sequence Number19187765
Product Code DQX
UDI-Device Identifier08717648124235
UDI-Public08717648124235
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K083706
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial
Report Date (Section B) 10/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2020
Device Catalogue Number1012068-09
Device Lot Number8070961
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/09/2018
Initial Report FDA Received Date10/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-